Luna® G3 BPAP System (Model LG3800-30VT)

Voluntary Recall – Targeted Pressure Data Reporting Correction – Originally Communicated October 8, 2026

Update Published: November 17, 2026

React Health (3B Medical, Inc.) is providing notice of a voluntary medical device correction (recall) involving specific Luna G3 BPAP System (Model LG3800-30VT) devices due to a data handling issue that may result in incorrect therapy pressure statistical data when the device is operated with Targeted VT enabled.

This notice provides important information for durable medical equipment (DME) providers, hospitals, healthcare facilities, clinicians, and patients using affected devices.

Explanation of the Issue

React Health is recalling specific units of the Luna G3 BPAP System (Model LG3800-30VT) due to a data storage and transmission error while Targeted VT is enabled. Devices operating with Targeted VT enabled will not record the actual pressure statistical data delivered to the patient on the internal SD card or transmit the pressure to the cloud-based management system (React Health Connect).

Users that have not enabled the Targeted VT feature on their Model LG3800-30VT device will not experience this issue.

Potential Risk and Clinical Considerations

The hazardous condition requires the device to be operated with Targeted VT enabled and monitored remotely via either data downloaded to the SD card or via wireless communication to the cloud-based management system (React Health Connect). The consequence of the inaccurate pressure data in historical therapy reports could lead a clinician to adjust device settings based on inaccurate therapy pressure data, which could result in delayed therapy until the setting is corrected. Design features that reduce the risk of this issue include the device adjusting pressure delivered to the patient within the clinician defined minimum and maximum pressures delivering a tidal volume equal to the clinician set tidal volume, within tolerance.

To date, React Health has received fifteen (15) complaints with two (2) serious adverse reports associated with this Luna G3 BPAP System (Model LG3800-30VT) Targeted VT enabled issue.

Clinicians should be aware that inaccurate historical pressure data may lead to misinterpretation of therapy performance, adjustments to device settings based on inaccurate information, or delays in recognizing opportunities to optimize patient therapy. Clinicians should consider these limitations when reviewing historical therapy reports and determining whether additional patient assessment or therapy adjustments are appropriate.

Actions for Patients/Users

Patients using a Luna G3 BPAP System (Model LG3800-30VT) Targeted VT enabled feature should do the following:

  1. Immediately contact your health care provider/physician and/or DME to assess and update prescribed therapy settings as needed.
  2. Do not independently change your prescribed therapy settings or discontinue respiratory therapy without guidance from your health care provider. Your healthcare provider can determine whether your device is affected, evaluate your therapy needs, and coordinate any necessary corrective action or alternative therapy support.
  3. Continue using the Luna G3 BPAP System as prescribed until you receive additional guidance from your health care provider.
  4. Complete the ‘Customer Recall Notification Response Form’ included in the React Health Recall Notification dated October 8, 2026, and send it to React Health within seven (7) days via e-mail to: 3B-Recall@reacthealth.com, for installing firmware correction.
  5. If you have obtained your Luna G3 BPAP System (Model LG3800-30VT) through a DME provider, contact your DME to install the firmware correction.

Actions for Healthcare Professionals and DME Providers

React Health requests that DME providers, hospitals, healthcare facilities, and physicians/clinicians review this information and follow the corrective actions outlined in the original customer recall notification dated October 8, 2026, including as below:

  1. Discontinue Targeted VT: Immediately discontinue use of Targeted VT if enabled on affected Luna G3 BPAP System (Model LG3800-30VT) devices. Healthcare providers should coordinate any necessary therapy transition to maintain appropriate patient support.
  2. Inspect affected inventory: Inspect current stock, quarantine or segregate unused affected products, and remove affected devices from clinical service until the corrective action is completed.
  3. Maintain patient therapy: Use an unaffected PAP device for patient support, as directed by the treating clinician.
  4. Review historical therapy data: Physicians/Clinicians should recognize that historical pressure data recorded while Targeted VT was enabled may not accurately reflect the actual pressures delivered during therapy. This problem should be considered when interpreting patients’ therapy reports, assessing patient response, or determining whether adjustments to prescribed therapy are appropriate and/or required.
  5. Notify affected recipients: If affected devices have been distributed or transferred to other facilities, caregivers, or end users, promptly notify them of this recall and provide a copy of the applicable recall notification.
  6. Complete the firmware correction: Coordinate with React Health or the DME provider to obtain and install the firmware correction in accordance with the instructions provided.

Firmware Correction and Customer Support

React Health has developed firmware version G3-2.00.80 to correct the identified pressure-data reporting issue. Firmware update SD cards are readily available through React Health, allowing affected devices to be updated without requiring permanent device replacement.

React Health is working directly with DME providers and healthcare facilities to coordinate distribution and implementation of the firmware correction. Upon receipt of the completed Customer Response Form, React Health will arrange shipment of the firmware update along with installation instructions and an updated confirmation letter.

This firmware update will allow the collection and display of reported pressure data of delivered therapy while Targeted VT is enabled. All historic data collected with Targeted VT enabled prior to this upgrade will not be changed and will reflect previously recorded non-fluctuating therapy pressures in device storage and the related cloud platform applications.

Affected customers should follow the firmware installation instructions, historical data management procedures, and update confirmation requirements provided in the original recall notification. Following installation of firmware version G3-2.00.80, the original recall instructions direct customers to erase historical usage data through the device's Clinical Menu – Maintenance – Erase Data function.

Additional Information

React Health has communicated recall instructions to affected customers and is coordinating the firmware correction to address the identified issue.

DME providers and healthcare facilities requiring assistance with affected devices, firmware updates, or recall instructions should contact React Health Customer Support at 863-226-6285, Monday through Friday, 8:00 a.m. to 5:00 p.m. Eastern Time.

If an adverse reaction or quality problem has been experienced with the use of this product, it may be reported using one of the following options:

Patients experiencing adverse reactions or quality problems with the use of this device also should report the issues directly to their DME. Alternatively, patients may report to React Health using the Customer Contact Form

React Health appreciates the cooperation of our healthcare partners in implementing this correction. Patient safety remains our highest priority, and we are committed to supporting our customers and healthcare professionals throughout the recall process.